.jpg)
The new Rigel 377 incorporates a range of custom settings that include a variety of test sequences and power distribution curves which make testing fast, easy and accurate. Utilising the large full color display, built-in memory and data management facilities, the Rigel 377 is the first fully stand-alone diathermy tester that can offer true user friendly operation and operator safety. The large array of internal resistors, ranging from 10-5100Ω in 10Ω steps provide not only the most accurate and detailed power curves, the Rigel 377 also advocates safe working practise by providing all necessary resistors within the enclosure.
The remote electrode monitoring test offers a unique isolated and electronically driven potentiometer to provide accurate and fast testing of alarms. In addition, the Bluetooth-enabled technology allows the wireless connectivity of the tester with PCs and other accessory equipment for fast and convenient downloading of performance data and the uploading of manufacturers' power curves and manufacturer specific test programs.
Other special features of the Rigel 377 include in-built graphics to display connection diagrams, program settings, power curves, alarm settings, data management and data analyses.
Custom Test Settings
The unique color user-interface provides both graphic as well as numeric representation of the test data. Showing manufacturer specific curves and allowable tolerances, the large color screen provides instant performance indication.
Use the graphic representation to quickly determine whether the test data meets the set criteria.
Use the numeric representation of the data to analyse parameters such as Peak voltage, crest factor, power and current at each resistance point etc.
Remote Electrode Monitoring
Use the rotary encoder to set high and low alarm points as well as controlling the potentiometer, scaling up or down in manual or automatic mode. Then, simply capture the alarm using the dedicated fast key on the screen.
Data will be stored onboard for future traceability.
HF Leakage
Individual settings can be selected using the simple configuration template. The settings are automatically translated into a detailed scheme as shown below;
Detailed instructions on the leakage path and connections between each setting are provided in a clear and color coded overview.
Color coded connections correspond to the matching connection on the tester.
Applications:
To verify the correct performance of Electrosurgical devices during:
a. Routine testing
b. Type testing
c. Production line testing
Unique training tool for testing on Electrosurgical devices
Specification
Power measurement True RMS value of applied waveform
Power rating 0 - 500W (RMS)
Duty cycle 100% 0-250W, 50% >250W
Accuracy ±1W or 5% of value
Voltage (peak) 0 - 10KV (Peak) - Closed load only
Accuracy ±10% of value.
Measurement is taken between the active and dispersive electrodes
with closed load only.
Voltage (RMS) 0 - 500V
Accuracy ±2V or 2% of value
Current 0 - 2200mA
Accuracy ±10mA or 2% of value
Crest factor 1.4 - 20 (Vpeak / V RMS)
The higher of the two peak voltage surements is used for calculation
RMS Bandwidth
Instrumentation only 30 Hz to 10 MHz (-3 dB)
With loads: 30 Hz to 2.5 MHz (-3 dB)
Variable loads 10 - 5100Ω, steps @ 10Ω (510 steps)
Accuracy 1% of R
Load array Ceramic Resistors (Non inductive)
Measurement delay Foot switch delay selectable between
200 - 5000mS (10mSec resolution)
RF Leakage (High Frequency Leakage)
Active From active part to earth
Passive From plate - receptacle - to earth
Load Variable see Power Measurement
Fixed 2 x 200Ω / 250 Watt each
Remote Electrode Monitoring
Range 0 - 500Ω, steps @ 0.5Ω steps
Motor driven potentiometer
Alarm register High and low, manual confirmation
Ranging Manual or automatic
Output Connectors
Remote foot switch control (CUT) 2 x 4 mm - yellow, Single relay contact
Remote foot switch control (COAG) 2 x 4 mm - blue, Single relay contact
High Frequency leakage 1 x 4 mm - black (common), 2 x 4 mm - red
Variable load output 2 x 4mm (for power measurement)
Red = Active, Black = Plate
Fixed load input 3 x 4 mm (3 x Blue) (200Ω / 500W)
USB PC Download / Future PC software
Oscilloscope output 5V/A, 100mA RF current minimum input
Isolation 10kV Isolation between measurement device and enclosure
Low Frequency Filter 100 Hz filter to avoid low-frequency disturbance or interference
Storage and Recall
Memory Approx 5,000 records
Output CSV format
Waveform
General Specifications
Dimensions 370 X 300 X 204mm
Weight 7 Kg
Operating temperature 15 °C to 35 °C
Storage temperature 0 °C to 50 °C
Mains power 115/230 VAC; 48 to 66 Hz, 35 VA