Compact Manual Electrical Safety Analyser
- IEC / EN 60601-1
- Compact and robust enclosure
- 1-25A Earthbond current
- Portable & Lightweight
- Full IEC 60601 Leakage Test
- Up to 25A AC Test Current
- Large LCD Display
- Separate AC & DC Value
![]()
![]()
![]()
![]()
![]()
![]()
The Rigel 266 Plus Biomedical Electrical Safety Analyzer is the most compact electrical safety analyzer capable of delivering 25 Ampere earthbond current and testing in accordance with IEC/EN 60601-1, MDA DB9801, and AS/NZ 3200.
The Rigel 266 Plus Biomedical Electrical Safety Analyzer boosts a large LCD screen and rotary selector and test routines can be manually controlled for quick and easy testing of medical devices. Single Fault Conditions can be manually introduced to minimize power breaks and reduce test times.
With its compact design and full IEC 60601-1 features, the Rigel 266 Plus Biomedical Electrical Safety Analyzer provides the user with a highly effective and portable test solution.
Rigel 266 Plus Biomedical Electrical Safety Analyzer
| AC Earth Bond Test | |
|---|---|
| Open Test Voltage | 6V rms nominal |
| Test Current | 1A, 10A, 25A AC |
| Range | 40mohms – 19.99ohms |
| Resolution | 0.01ohms |
| Accuracy | +/- 5% of reading, +/- 2 counts |
| Measurement | 4-wire measurement, floating earth |
| Insulation Test | |
|---|---|
| Test Voltage | 500V d.c. nominal (0.5Mohms load) |
| Range | 100kohms - 100 Mohms |
| Resolution | 0.01 Mohms (<100 Mohms) |
| Accuracy | +/- 5% of reading, +/- 2 counts (100kohms - 20Mohms) |
| All Leakage Tests | |
|---|---|
| Input Impedance | >1Mohm |
| Frequency Response | as per IEC 60601-1 requirements |
| Accuracy | +/- 5% of reading, +/- 10 counts |
| Range | |
|---|---|
| Earth Leakage | 0.04mA – 9.999mA |
| Enclosure Leakage | 0.04mA – 9.999mA |
| Patient Leakage | 0.004mA – 9.999mA |
| Patient Auxiliary Leakage | 0.004mA – 9.999mA |
| Patient SIP / SOP Leakage | 0.04mA – 9.999mA |
| Patient F-Type Leakage | 0.025mA – 9.999mA |
| F-Type Open Circuit Voltage | 110% +/-20% of Mains Input Voltage |
| Load Test | |
|---|---|
| Measured Load | 0.06 - 4kVA |
| Measured Voltage | Mains Supply +/-10% |
| IEC Lead Test | |
|---|---|
| Test | 40V a.c., 1mA nominal |
| Detects | Good, Open, Short, Reverse |
| Mechanical | |
|---|---|
| Size | 310mm x 210mm x 80mm |
| Weight | 4kg |
| Environmental | |
|---|---|
| Operating | 0ºC to 40ºC (non condensing) |
| Storage | -10ºC to 50ºC (non condensing) |
| Maximum R.H | 90% |
| Part Numbers | |
|---|---|
| 350A910 | UK (13A) |
| 350A913 | Schuko (16A) |
| 350A915 | Danish (10A) |
| 350A917 | French (16A) |
| 350A918 | Czech / Polish (16A) |