Universal Automatic Biomedical Electrical Safety Analyzer
- IEC / EN 60601-1 / 61010
- On-board printer and memory
- 0.1 to 25A Earthbond current
![]()
![]()
![]()
![]()
![]()
![]()
The Rigel 277 Plus Biomedical Electrical Safety Analyzer is a portable medical electrical safety analyzer combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment) including Touch Leakage, Voltage Measurement and dedicated IEC 61010 Measuring Device (Body Model).
Customized test routines can be created to meet local standards and variants of safety test standards. The Rigel 277 Plus Biomedical Electrical Safety Analyzer performs manual, automatic and semi automatic test routines allowing full control of power up and power down sequence, thus reducing the overall test times.
The Rigel 277 plus boosts a large full graphics display, internal memory, QWERTY keyboard and printer making this a completely stand alone biomedical electrical safety analyzer.
Rigel 277 Plus Biomedical Electrical Safety Analyzer
| AC Earth Bond Test | |
|---|---|
| Test Current | 100mA, 1A, 10A, 25A |
| Current Source | Regulated |
| Open Test Voltage | 6V rms nominal |
| Range | 0.01ohms - 19.99ohms / 0.01ohms counts |
| Accuracy | ± 5% of reading, ± 2 counts |
| Measurement | 4-wire, floating earth |
| Insulation Test | |
|---|---|
| Test Voltage | 350 / 500V d.c. nominal |
| Auto Ranging | 100kohms - 20 Mohms (L) / 0.01Mohms counts 20Mohms – 100Mohms (H) / 0.01Mohms counts |
| Accuracy | ± 5% of reading, ± 2 counts (L) ±10% of reading, ± 2 counts (H) |
| Open Circuit Voltage | <750VDC |
| IEC 60601 Leakage Test | |
|---|---|
| Supply Voltage | Incoming mains (230/110VAC) |
| F-Type Open Circuit Voltage | 110% +/-20% of Mains Input Voltage |
| Maximum Load | 16A (13A for UK model) |
| Input Impedance | 1Kohm – as per 60601 MD |
| Frequency Response | as per IEC 60601-1 requirements |
| Range | 0.000mA – 9.999mA / 0.001 counts |
| Accuracy | ± 3% of reading, ± 4 counts |
| IEC 61010 Leakage Test | |
|---|---|
| Supply Voltage | Incoming mains (230/110VAC) |
| Maximum Load | 16A (13A for UK model) |
| Input Impedance | 2Kohms – as per 61010 MD |
| Frequency Response | DC – 2.5kHz (as per IEC 61010) |
| Range | 0.10mA – 3.50 mA / 0.01 counts |
| Accuracy | ± 3% of reading, ± 4 counts |
| Load Test | |
|---|---|
| Measured Load | 0.06 - 4kVA |
| Measured Voltage | Mains Supply +/-10% |
| IEC Lead Test | |
|---|---|
| Test | 40V a.c., 1mA nominal |
| Detects | Good, Open, Short, Reverse |
| General Specification | |
|---|---|
| Size | 400mm x 320mm x 160 mm |
| Weight | Appropriately 9 kg |
| Operating | 0ºC to 40ºC (non condensing) |
| Storage | -10ºC to 50ºC (non condensing) |
| Maximum R.H | 90% |
| Part Numbers | |
|---|---|
| 350A910 | UK / Ireland |
| 350A913 | Euro - Schuko |
| 350A917 | French |
| 350A915 | Danish |
| 350A918 | Czech / Polish |
| 350A914 | US (110V – 60Hz) |
| 350A912 | Australian |