Rigel Uni-Therm

Electrosurgical Analyzer

  • IEC 60601-2-2 compliant
  • Power distribution from 0-5115Ω
  • HF leakage test
  • Return plate security test (Rem)
Rigel Uni-Therm Electrosurgical Analyzer

Rigel Uni-Therm Electrosurgical Analyzer

Click
to Request a Quote

The Rigel Uni-Therm is a high performance electrosurgical analyzer with the emphasis on speed and ease of testing all modern day electrosurgical generators.

Utilizing a high resolution, low-inductance load bank the Rigel Uni-Therm electrosurgical analyzer can measure high frequency leakage and power distribution from 0Ω to 5115Ω in 5Ω increments while problems with the patient plate can be simulated from 1-475Ω in 1Ω increments.

Boasting a large color screen, the Rigel Uni-Therm electrosurgical analyzer displays connection diagrams, program settings, power curves, alarm settings, data management and data analysis.

Featuring Bluetooth and USB connectivity, test data can be stored within the instrument and transferred wirelessly to Rigel’s Med-eBase asset management software.

Watch a video about the Uni-Therm electrosurgical analyzer - North America:

 

Watch a video about the Uni-Therm electrosurgical analyzer - Latin American Spanish:

 

Watch this "how to" video on accomplishing a preventive maintenance (PM) of ForceTriad using the Rigel Medical Uni-Therm analyzer

 

Watch this "how to" video on accomplishing a preventive maintenance (PM) of a Covidien, ValleyLab, Force FX ESU using the Rigel Medical Uni-Therm analyzer:

Watch this "how to" video on accomplishing a preventive maintenance (PM) of a MEGAPOWER surgical generator, manufactured by MEGADYNE using the Rigel Medical Uni-Therm analyzer:

Rigel Uni-Therm Electrosurgical Analyzer

Click
to Request a Quote

Technical Specification:

Specifications
Power measurement True RMS value of applied waveform
Power rating 0 - 500W (RMS)
Accuracy ±(1W + 5% of value)
Duty cycle 100% up to 60 seconds
Load bank 0 - 5115Ω
Resolution
Voltage (peak) 0 - 10kV (Peak) - Closed load only
Accuracy ±(10% of reading +15v).
Measurement is taken between the active and dispersive electrodes with closed load only.
Voltage 0 - 700V (RMS)
Accuracy (5V + 10% of value)
Current 0 - 6000mA (RMS)
Accuracy ±(2% of reading +10mA)
Crest factor 1.4 - 20 (Vpeak / V RMS)
The higher of the two peak voltage measurements is used for calculation
RMS Bandwidth
Instrumentation only 30 Hz to 10 MHz (-3 dB)
With loads: 30 Hz to 2.5 MHz (-3 dB)
Variable loads 0 - 5115Ω, steps @ 5Ω (1023 steps)
Accuracy 1% of R
Load array Ceramic resistors (Non inductive)
Measurement delay Foot switch delay selectable between 200 - 5000ms (10mSec resolution)
RF Leakage (High Frequency Leakage)
Active From active part to earth
Passive From plate - receptacle - to earth
Load Variable, see power measurement and additional fixed 2 x 200Ω for bipolar setup
Return Electrode Monitoring
Range 1- 475Ω, in 1Ω steps by motor driven potentiometer
Alarm register High and low, manual confirmation
Ranging Manual or automatic
Output Connectors
Remote foot switch control (CUT) 2 x 4mm - yellow, single relay contact
Remote foot switch control (COAG) 2 x 4mm - blue, single relay contact
High Frequency leakage and power measurements Through 4mm sockets
   
USB PC download
Oscilloscope output 5V/A, 100mA RF current minimum input
Isolation
10kV Isolation between measurement device and enclosure
Low Frequency Filter
100 Hz filter to avoid low-frequency disturbance or interference
Storage and Recall
Memory Approx 5,000 records
Output CSV and SSS format
General Specifications
Dimensions 370 X 300 X 204mm - 14.6 X 11.8 X 8 Inches
Weight 10 kg / 22 Lbs
Operating temperature 15 °C to 40 °C
Storage temperature 0 °C to 50 °C
Mains power 120/230 VAC +/-10%; 48 to 66 Hz, 35 VA
Part number 398A912
Service & Warranty
Uni-Therm comes with a free upgraded 24 month warranty subject to terms and conditions.
A range of Med-eCare plans are also available.