NIBP / SPO2 and ECG Simulator
- Accurate 6 in 1 vital signs simulation
- Hand-held and Battery powered
- On-board memory and keyboard
![]()
![]()
![]()
![]()
![]()
![]()
The Rigel UNI-SiM is the world’s first battery powered 6 in 1 vital signs simulator able to undertake six synchronized vital signs parameters tests.
Enabling quick, easy and accurate simulations of NiBP, SpO2, ECG, temperature, IBP and respiration functionality, the Rigel Uni-Sim is one of the most valuable tools around today.
Utilizing new and advanced technologies in both SPO2 and NIBP simulation, reduces errors and minimizes the test time. Manufacturer specific simulation curves can be created and uploaded to further aid the accuracy.
Bluetooth connectivity to a range of accessories and Med-eBase, comprehensive database software, ensures fast and easy download of test results, managing your asset database, creation of test sequences and producing test certificates.
Rigel UNI-SIM Vital Signs Simulator
| Non-invasive Blood Pressure | |
|---|---|
| Simulation Waveform | Oscillometric |
| Integrated Pump | 0 to 350 mmHg user configurable |
| Leak Test | User configurable between 0-350mmHg |
| Chronometer | Configurable upto 999 secs |
| Digital Manometer | 0 - 410 mmHg |
| Pressure Accuracy | +/- 0.5% FS |
| Pressure Units | mmHg, inHg, kg/cm2, cmH2O, mBar, PSI, inH2O and kPa |
| Oxygen Saturation | |
|---|---|
| Range | 50 to 100% |
| Accuracy | ± 1 bpm |
| Chronometer | Test option to test response time SpO2 monitor |
| ECG Arrhythmia Simulator | |
|---|---|
| ECG | Full 12 lead simulation including high level output |
| Wave Forms | Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T |
| Rate | 20 – 300 BPM typed value |
| Performance Waveforms | Sine, Square, Triangle, and Pulse |
| Pacer Waveforms | Synchronous Atrial, Asynchronous Atrial |
| Pacer Only | Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial & Ventricular Pacer Heart rate 70 BPM, R Wave Detection |
| Arrhythmia Waveforms | |
|---|---|
| Atrial | Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Contraction |
| Atrial Conduction | First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock |
| Ventricular | Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction – Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Focal PVC |
| Temperature | |
|---|---|
| Simulation | YSI 400 / 700 Static |
| Range | Preset at 25, 33, 37and 41°C |
| Respiration | |
|---|---|
| Rates | 5, 10, 15, 30, 60, 120, 180 Breaths per Minute |
| Base Resistances | 250, 500, 750, 1000 ohms |
| Apnoea Simulation | 0 – 60 seconds duration 0 – 300 seconds interval |
| Invasive Blood Pressure | |
|---|---|
| Channels | 2 channels |
| Static | 0 to 300mmHg. Typed values |
| Dynamic | 0 - 300mmHg for systolic & diastolic |
| General Specification | |
|---|---|
| Operation | Battery cell, insitu charge |
| Battery Charger | 110-230VAC, 50/60 Hz |
| Battery Life | 4-24 hours testing depending on simulations |
| Memory Capacity | Approximately 10,000 records |
| Communication | Via Blue Tooth |
| Display | Monochorme, ¼ VGA full graphics |
| Keypad | Alpha-numeric |
| Weight | <1.5 kg |
| Size (L x W x D) | 270 x 110 x 75 mm / 10.5 x 4 x 3” |
| Operating Conditions | 10-30°C, 0-90% RH - NC |
| Storage Environment | -15°C - +60°CC |
| Environmental Protection | IP 40 |
| Part Number | |
|---|---|
| 370A930 | |