The World's Most Advanced Portable Electrical Safety Tester
Battery Powered. Full Alpha-numeric Keyboard
A battery powered tester means that you spend less time plugging and unplugging your tester and more time testing, and gives you the flexibility to test when and where it is right for you.
Blue Tooth Wireless Networking
Free yourself of the cables and leads you need to run your accessories and free your time to carry out more testing.
Hand-Held. Smaller and easier to use.
Move faster before, during and after tests and inspections, with this easy to handle and use ‘comprehensive’ tester and data logger.
PrimeTest 300 is 'The Complete Safety Tester'Test the Device. Test the Circuit. Record asset locations and non-electrical test results.
Ground Continuity
Output Current (Load 2W): +/- 200mA DC
Test Voltage (o/c): > 4VDC
Test Duration: Programmable within a maximum limit
Measuring Range: 0.01 - 19.99W
Operating Error: +/-5% + 2 counts
Measurement Resolution: 0.01W
Pass Value: User defined
Insulation Test
Test Voltages: 500VDC @ 1mA nominal
Test current: >1mA into 500kW; < 2mA into s/c
Test Duration: Programmable within a maximum limit
Measurement Range: 0.10 MW - 40MW
Operating Error: +/-5% +2 digits
Measurement Resolution: 0.01MW (Max)
Pass Value: User defined
Differential Leakage
Test Voltage: 120V +10%, -15%
Maximum Load current: 13 A rms
Test Duration: Programmable within a maximum limit
Measurement Range: 0.25 - 19.99mA
Measurement Resolution: 0.01mA
Operating Error: +/-5% of reading +/- 2 digits
Pass Value: User defined
Touch Leakage
Test Voltage: 120V +10%, -15%
Maximum load current: 13Arms
Test Duration: Programmable within a maximum limit
Measurement Range: 0.10 - 3.50mA
Measurement Resolution: 0.01mA
Operating Error: +/- 10% of reading +/- 2 digits
Pass Value: User defined
GFCI Test
Test Voltage: 120V +10%, -15%
Test Current: 30 mA rms sinusoidal
Test Current Accuracy: +10, -0%
Test Current Phase: 0°/180° auto switched on alternate
Test Duration: 2000 ms maximum if GFCI does not trip
Measuring Range: 10ms - 2000ms
Operating Error: ±10%
Measurement Resolution: 1ms
Ground Loop Resistance Test
Test Voltage: 120V +10%, -15%
Test Current: Such that it will not trip a 30mA GFCI
Measuring Range: 1W - 199W
Operating Error: ±10% +2 digits
Measurement Resolution: 0.1W (Max)
Lead Test
Test voltage: 9 VDC
Test Duration: 2s
Test: Live / Neutral Checks for o/c, s/c and crossed
Power Socket Test
Input voltage range: 120V +10%, -15% AC
Indicators: phase to ground voltage
phase to neutral voltage
neutral to ground