Complete Electrical Medical Safety Analyzer
The Rigel 277 Plus is a portable medical electrical safety analyzer combining IEC/EN 60601-1 compliance with additional test facilities for IEC/EN 61010 (Laboratory Equipment) including Touch Leakage, Voltage Measurement and dedicated IEC 61010 Measuring Device (Body Model).
In addition, test routines are preset for MDA DB9801and VDE 0751. Customized test routines can be created to meet local standards and variants of safety test standards such as AS/NZ 3200 and previous editions of IEC / EN 60601 standards.
The Rigel 277 Plus performs manual, automatic and semi automatic test routines allowing full control of power up and power down delay times. The combination of a full graphics display, large internal memory and integral qwerty keyboard provides a complete stand alone test system to confirm the electrical safety of both medical and non medical electrical equipment.
The Rigel 277 Plus is designed to increase the efficiency of test routines and to reduce the overall test time.
Test current - 100mA, 1A, 10A, 25A AC C.C.
Max test voltage - 6V rms O/C
Range - 0 – 19.99 Ohm
Accuracy - +/- 5% +/- 2 counts
Insulation measurements (Enclosure, Patient, IEC lead)
Voltage - 350V / 500 VDC
Range, Accuracy - 100K-20M Ohm, +/- 5% +/- 2 counts
20M – 100M Ohm+/- 10% +/- 2 counts
Max O/C voltage - <750V
Leakage measurements (Ground, Enclosure, Patient, Auxiliary, F-type, Substitute)
Input impedance - >1M Ohm
Frequency response - DC to 1 MHz ( -3 dB )
Range - 0 – 9.999 mA
Resolution - 1uA
Accuracy - +/- 5% +/- 2 counts
Power measurement - 0.1KVA – 4 KVA, 50-60Hz
General
Supply voltage - 99-121 V / 207 – 253 V 50/60Hz
Weight ~ 20 lbs.
Dimensions - 15.8" x 12.6" x 6.3"