Rigel UNI-SIM Lite

Vital Signs Simulator (NIBP / SPO2 and ECG)

  • Compact and cost-effective
  • Simultaneous vital signs simulation
  • Fast start up and single button simulation
  • User definable systolic and NiBP simulations
  • Universal SpO2 simulation
  • Easy and accurate probe placement with PULS-R
Rigel UNI-SIM Lite Vital Signs Simulator

Rigel UNI-SIM Lite Vital Signs Simulator

Click
to Request a Quote

The UNI-SiM Lite is a handheld and battery-operated vital signs simulator, designed to be a cost effective and portable solution for carrying out spot checks and full PM’s on vital signs monitors, wherever the job takes you.

With synchronized vital signs parameters, fast boot up and single button simulation to repeat the last simulation value in seconds, the UNI-SiM Lite is a quick and easy-to-use simulator for basic spot checks.

User configurable heart rate, SpO2 values, systolic and diastolic NiBP pressures make the UNI-SiM Lite a truly versatile tool capable of meeting even the most demanding test protocols.

Compatibility with the Rigel PULS-R universal SpO2 simulation finger enables accurate and repeatable SpO2 values. Unique probe positioning LEDS help to ensure the correct positioning of SpO2 probes, improving accuracy and repeatability

Rigel Uni-Sim Lite with PULS-RThe Rigel PULS-R SpO2 Finger Simulator can also be used with the UNI-SiM Lite vital signs simulator.

Able to simulate accurate SpO2 simulations in 1% resolution from as low as 30%* using the pre-programmed manufacturer specific R-curves,the PULS-R can also be configured to meet customer specific R-curves using the simple to use configuration tool.
* subject to monitor capability

 

 

Free Vital Signs Booklet

We have created an in-depth guide to measuring and simulating vital signs.
 


Download now

Rigel UNI-SIM Lite Vital Signs Simulator

Click
to Request a Quote

Technical Specification:

Non-invasive Blood Pressure Simulation
Waveform Oscillometric
Pulse Volume High, Medium, Low, Paediatric
Heart Rate 20 - 300BPM
Integrated Pump 0 to 350 mmHg user configurable
Leak Test User configurable between 0-350mmHg
Chronometer Configurable up to 999 secs
Digital Manometer 0 - 410 mmHg
Pressure Accuracy +/- 0.5% FS
Pressure Units mmHg, inHg, kg/cm2, cmH2O, mBar, PSI, inH2O and kPa
Oxygen Saturation
Range 30 to 100%
Repeatability ± 5%** of reading between 30-59% SpO2
± 3% of reading between 60-99% SpO2
± 3% of reading between 90-100% SpO2
Accuracy of simulation when used with the corresponding R-curves
*Based on using the same probe and monitor setup
**Note that some monitor types might not be able to display low range sats
Heart Rate 30-300BPM***
Accuracy ± 1 bpm
Compatibility Beijing Choice
Criticare
GE Tuffsat
Masimo
Mindray
Nellcor
Nellcor Oximax
Nihon Kohden
Nonin
Novametrix
Philips / HP
***Subject to monitor capability
ECG Arrhythmia Simulator
Simulation 5 lead simulation including high level output on Normal Sinus Rhythm (NSR)
ST Elevation ST Depression, Myocardial Infarcation, Tall T
Heart Rate 20 – 300BPM
Accuracy ±1BPM
Amplitudes 0.5 / 1 / 1.5 / 2 / 2.5 / 3 / 3.5 / 4 / 4.5 / 5mV
Accuracy ±2%
Connection High-Level ECG 3.5mm jack plug
ST Elevation / Depression
Heart Rate 20 – 300BPM
Elevation % 7%, 13%, 20%
Elevation Slope Positive, Negative, Flat
Myocardial Infarction
Type Ischemia, Injury, Infarction, Inferior Infarction
Heart Rate 20 – 300BPM
Tall T
Heart Rate 80BPM
T Wave Amplitude 0 – 1.2mV (steps of 0.1mV)
Arrhythmia Waveforms (Atrial)
Simulation Full 12 lead simulation
Amplitudes 0.5 / 1 / 1.5 / 2 / 2.5 / 3 / 3.5 / 4 / 4.5 / 5mV
Heart Rate (where applicable) 20 – 300BPM
Atrial Conduction
Sinus Arrhythmia (SA), Missing Beat, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Contraction
Atrial Conduction First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock
Atrial
First Degree AV Block, Second Degree AV Block - Mobitz I, Second Degree AV Block - Mobitz II, Third Degree AV Block, Right Bundle Branch Block (RBB), Left Bundle Branch Block (LBB), Left Anterior Hemiblock
Ventricular
Premature Ventricular Contraction - Intermittent Premature Ventricular Contraction – Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation Fine (VFBF), Ventricular Fibrillation Coarse (VFBC), Monomorphic Ventricular Tachycardia (MVT), Polymorphic Ventricular Tachycardia (PVT), Right Focal (PVC)
Performance Waveforms
Shape Sine, Square, Triangle and Pulse
Rates 0.1 to 0.9Hz (in steps of 0.1)
1 to 100Hz (in steps of 1)
Amplitudes 0.5 / 1 / 1.5 / 2 / 2.5 / 3 / 3.5 / 4 / 4.5 / 5mV
Pulse 1mV, 4 sec delay (20ms pulse duration)
Pacer Waveforms
Available Synchronous Atrial, Asynchronous Atrial, Paver only, Ventricular Pacer, Atrial & Ventricular Acer
QRS 1mV
Pacer Pulse Amplitude 0.1 – 2mV
Pacer Pulse Polarity Positive, Negative
Pacer Pulse width 0.1 – 2ms
R Wave Detection
Heart Rate 70BPM
R Wave Width 10 – 120ms (steps of 10ms)
Temperature Simulation
Simulation YSI 400 / 700 Static
Range Preset at 25, 33, 37 and 41°C
Accuracy ±0.1°C
Default Setting YSI 400 37°C
Respiration Simulation
Rates 5, 10, 15, 30, 60, 120, 180 Breaths per Minute
Base Resistances 250, 500, 750, 1000Ω
Accuracy ±5%
Resistance Variations 0.1, 0.5, 1.0, 1.5Ω
Accuracy ±10%
Default Settings 15 BPM / 250Ω / 0.1Ω
Apnoea Simulation 0 – 60 seconds duration 0 – 300 seconds interval
Invasive Blood Pressure Simulation
Channels 2 channels
Static 0 to 300mmHg (0-5.8 PSI)
Dynamic 0 - 300mmHg for systolic & diastolic
Accuracy ± 1mmHg
Excitation Voltage 2V to 16V
Impedance 350Ω Nominal
Simulated Sensitivity 5μV/V/mmHg

General Specifications:

Operation Battery cell, in-situ charge
Battery Charger 100-240VAC, 50/60 Hz
Supply 12VDC centre positive
Battery Life 8 hours standby or a maximum of 200 NiBP simulations
Communication Via Blue Tooth
Display Monochorme, ¼VGA full graphics
Keypad Alpha-numeric
Weight <3.5lbs
Size (L x W x D) 10.5 x 4 x 3”
Operating Conditions 50⁰F-86⁰F, 0-90% RH - NC
Storage Environment 5⁰F - +140⁰F
Environmental Protection IP 40
Service & Warranty
UNI-SiM Lite comes with a free upgraded 24 month warranty (subject to terms and conditions, available at www.rigelmedical.com/registerproduct)
Standard Accessories (supplied with UNI-SiM Lite)
Carry case
ECG snap-on adaptors
NiBP tubing kit
Quick start guide
ECG adaptor module
 US power supply 120VAC
Optional Accessories
IBP connect cables
Temperature connect cables

Specifications - Puls-R:

Supported Default R Curves
Beijing Choice
Criticare
GE Tuffsat
Masimo
Mindray
Nellcor
Nellcor Oximax
Nihon Kohden
Nonin
Novametrix
Philips / HP
Heart Rate Setting 30-300BPM (subject to monitor compatibility)
Operating Temperature 0-104°F
Dimensions 3.08” x 2.68” x 1.14”
Finger Shape 1.58” x 0.71”
Part Numbers
UNI-SiM Lite
Puls-r
370A934
399A910
 